Testosterone
Analyt: | Testosterone |
Matrix: | Serum |
Measuring Method: | GC-ID/MS |
Standard: | Testosterone NMIJ CRM 6002-a |
Internal Standard: | ¹³C₂-Testosterone |
Sample Processing: | Samples are spiked gravimetrically with internal standard and equilibrated, then precipitated with aqueous KOH, centrifuged and the supernatant extracted into dichloromethane. Derivatisation is with cyclohexane-HFBA and subsequent extraction into cyclohexane. |
Measurement: | Measurement at m/z 680 and m/z 682 |